Transfer functions identification in the artery-tissues-hydrocuff system for the purpose of forming individual input parameters for artificial neural networks

Transfer functions identification in the artery-tissues-hydrocuff system for the purpose of forming individual input parameters for artificial neural networks 150 150 IEEE Open Journal of Engineering in Medicine and Biology (OJEMB)

Abstract:

Objective. The identification problem of dynamic processes occurring in the artery-tissue-hydrocuff system based on the Volterra series usage is considered. The identification procedures use algorithms based on the approximation of the required characteristics represented by segments of expansions by orthogonal Chebyshev polynomials. Results. During determining the expansion coefficients, a collocation …