Paul Neubauer

Associated articles

TBME, Featured Articles
Modeling of Interstitial Ultrasound Ablation for Continuous Applicator Rotation with MR Validation
Predictive modeling can play an important role in thermal ablative cancer treatments where estimated energy deposition may integrate into pre-operative planning software or combine with closed loop control. In this work, we develop a finite element model capable of simulating continuous rotation by a directional interstitial ultrasound applicator. The developed model simulates temperature maps for different rotation trajectories. The developed simulation was validated using magnetic resonance thermal imaging and demonstrated performance suitable for real-time control of robotically actuated conformal tumor ablation... Read more